Wxx02 (2000 N (450lbs) Version) |
Wxx04 (4000 N (900 lbs) Version) |
Input voltage [VDC] |
12 |
24 |
12 |
24 |
Static load (Fx), maximum [N (lbs)] |
2000 (450) |
4000 (900) |
Dynamic load (Fx), maximum [N (lbs)] |
2000 (450) |
4000 (900) |
Stroke lengths, standard [cm (inch)] |
10 (3.94), 20 (7.87), 30 (11.81), 40 (15.75), 50 (19.69) |
Speed version |
58A |
54A |
58A |
54A |
58A |
58A |
Speed, no load [mm/s (inch/s)] |
5.8 (0.23) |
11.0 (0.43) |
5.8 (0.23) |
11.0 (0.43) |
5.8 (0.23) |
5.8 (0.23) |
Speed, rated load [mm/s (inch/s)] |
4.0 (0.16) |
8.0 (0.31) |
4.0 (0.16) |
8.0 (0.31) |
4.0 (0.16) |
4.0 (0.16) |
Current draw @ rated max. load [A] |
4.5 |
9.0 |
2.2 |
4.5 |
9.0 |
4.5 |
Current draw @ stall/in-rush [A] |
14.0 |
21.0 |
8.0 |
11.0 |
21.0 |
11.0 |
End play, maximum [mm (inch)] |
0.5 (0.020) |
Operating temperature limits [°C (°F)] |
-25 to +40 (-13 to +104) (1) |
Duty cycle, maximum [%] |
10 |
On-time, maximum [s] |
180 |
90 |
180 |
90 |
90 |
90 |
Life, average [cycles] |
10,000 |
Sound level [dBa] |
< 45 |
Lead screw type |
acme |
Protection class |
IP67 |
Anti rotation mechanism(2) |
yes |
Analog feedback output signal voltage [VDC] |
0.5 - 4.5 |
Analog feedback output signal linearity [%] |
0.5 |
Digital feedback input voltage [VDC] |
3.8 - 24 |
Digital feedback output resolution [pulses/mm] |
10 |
5 |
10 |
5 |
10 |
10 |
Compliances |
IEC 60601-1, CSA C22.2, AAMI ES60601-1 (3) |
- Electronic limit switches (ELS) (4)
- ELS + low level switching (1) (5)
- Analog or digital position feedback, with or without ELS
- Mounting adapters turned 90°
|
? 1) Thrust is reduced at lower temperatures for 2000 N (450 lbs) high speed (54A) and 4000 N (900 lbs) units with ELS + low level switching (option E).
2) Prevents the extension tube from rotating if it is not fixed in the end.
3) 4 kN units greater than 30 cm of stroke are not ETL recognized. ETL certification available on N, E, P and D feedback options only.
4) Shuts off power at the end of stroke and all along the stroke at overload conditions. ELS is normally set for 120% of the rated dynamic maximum load.
5) Low level switching allows low level signals to control the actuator movement instead of having to switch the high amperage motor voltage.
|